Trigger selection
Triggering is the most co mplicated operation to perform for the
oscill
osco
pe because this instrument has many requirements that
have to be incidentally applied. and it requ ires an exact
synchronization as well.
CD
Trigger mode selection
Auto trigger mode :
Since the synchronized sweep always
appear
s. even though there
exists no signal nor has the trigger adjusting been done
improperly granting that there exists a signal. you have
nothing to worry about that erro rs can possibly arise from the
N
ORM
.
However. where the signal frequency is less than 25Hz. the
AUTO
cannot
be used. At this time. the measurement has to be
done
at
the NORM positi
on
.
NO RM trigger mode :
CRT b
eam
appe
ar
s only when the signal is synchronized.
This
trigger MODE
doc
s not cause the trace to appear in case that
there is no signal, that the synchroniza tion adjusting has b
een
d
one
improperly
and
that the venieal POSITION has been
incorrectly adj usted or that the YOLT
/D
IY switch has been
improperly positi
oned
.
TV-V. TV-H trigger m
ode
:
The
cleanly synchronized waveform c
an
be observed with
separating the waveform such as a co
mpo
site video signal
(Fig. 2-7(a» into the horizontal
and
vertical
compon
ents by
adding
a TV sync separat
ion
circuit,
Fo
r the sync
hron
ization
of
vertical
compo
nents of the TV signal (Fig. 2-7(b» . set the
Trigger M
OD
E switch switch to TV-Y. For the sync
hro
nization
of
horizontal components of the TV signal
(Fig
. 2-7(c
».
set the
Trigger MODE switch to TV-H. When the TRI
GGE
R
has
been
separated (Fig. 2-7
(d
».
the TV sync polarity should be
negative/ - ).